EasyManuals Logo

CEM Liberty Blue User Manual

CEM Liberty Blue
63 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #48 background imageLoading...
Page #48 background image
44
Maintenance of the Liberty Blue
NOTE
As each position is backushed, the Last Backush date will be updated to the current date and time.
8. Once all reagent positions have been backushed, the manifold will be cleared, and the system will perform a
backpurge of selected positions.
9. After backushing, observe the volume in the amino acid bottles. If the bottles have less than 2 mL of Main Wash
added, replace the dip tube lter in the bottle(s) that are low and perform a backush on those specic positions
again. If volume is still below 2 mL, contact CEM.
10. Once all positions have been backpurged, remove the bottles from those positions and dispose of the liquid.
Backflush Deprotection
1. Select “Maintenance” from the “Options” menu.
2. Select the Backush/Backpurge folder and then select the Backush Deprotection operation
3. Click Run Operation.
4. The software will prompt to verify the reaction vessel is empty. Ensure the reaction vessel is empty, then click
OK to continue.
5. Solvent will be added to the reaction vessel, then backpurged into the deprotection bottle. To avoid lling
the lines with the solvent, the software will prompt to replace the bottle. Replace the bottle, then click OK to
complete the backush operation.
Timed Addition Calibration
There are four timed additions that must be calibrated for optimal functionality of the Liberty Blue: Wash, Wash
Thru Manifold, Deprotection, and Wash Thru RV Bottom. To calibrate a timed addition:
1. Place the reaction vessel into the volume calibration stand.
1.1. Remove the reaction vessel assembly from the cavity of the Discover.
1.2. Insert the reaction vessel into the calibration stand so that the drain line from the bottom of the
vessel ts into the slot on the right of the stand.
2. From the Calibration screen, under the Timed Addition folder, select the addition to be calibrated. The table
on the right will update with the calibration history for the selected addition.
3. Click the Calibrate button to begin the calibration operation.

Table of Contents

Other manuals for CEM Liberty Blue

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the CEM Liberty Blue and is the answer not in the manual?

CEM Liberty Blue Specifications

General IconGeneral
BrandCEM
ModelLiberty Blue
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals