46
Maintenance of the Liberty Blue
4. The software should prompt a Waste Full error. In addition, the Waste indicator will turn red.
5. If the Waste Full error does not prompt, ensure the 9-pin connector is securely plugged into the Liberty Module.
If the plug is secure and the error does not prompt, replace the waste sensor assembly.
6. Release the oat sensor, connect the cap to the waste container, then click OK to clear the Waste Full error.
Standby Procedure
If the Liberty Blue will be idle for a period of two weeks or more:
1. Remove all reagents (amino acids, activator, activator base) from the system.
2. Select “Maintenance” from the “Options” menu.
3. Use the Change Bottle Deprotection operation to replace the deprotection bottle with an empty bottle and
properly clear the lines.
NOTE
During the priming step, the system will get an Empty Deprotection Bottle Error. This is normal. Click
Cancel to clear the error and end the Change Bottle operation.
4. Backush all positions, including deprotection.
5. Use the Change Bottle Main Solvent operation to replace the main was solvent bottle with an empty bottle and
properly clear the lines.
NOTE
During the priming step, the system will get an Empty Solvent Bottle Error. This is normal. Click Cancel to
clear the error and end the Change Bottle operation.
6. Use the Depressurize System operation to depressurize the system.
Prior to using the Liberty Blue after it has been in standby:
1. Connect the main wash solvent using the Change Bottle Main Solvent operation to properly prime the lines and
pressurize the bottles.
2. Connect the deprotect bottle using the Change Bottle Deprotection operation to properly prime the lines and
pressurize the bottle.
3. Perform a backush on all positions.
4. Perform a Timed Addition Calibration for all timed additions.