CF Control Installation
07/2017
Page
15
ATTENTION
:
Failure to comply with the assembly instructions can result in
measurement errors.
The water sampling is to be arranged in such a way that representative sample water
is available at all times.
a) Extraction from the pool through suitable borings in the pool wall, about 30-50
cm below the water surface. (Ideal situation!)
b) Extraction from the pressure side of the circulation pump, in front of the filter. It
must be ensured in this case that the sample water is not altered/manipulated
by fresh water. If required, the sample water is to be taken from the floor
suction circulation.
c) Extraction from the overflow gutter.
WARNING: Check that water is overflowing at all times!
If the sample water pressure is less than 0.1 bar, or the sample water line is
exceeding a length of 10 m, a sample water pump must be installed.
WARNING: A fiber filter is installed in the sample water line to prevent the
measuring cell from getting clogged by dirt. The fiber filter must be
cleaned at regular intervals (every 1-2 weeks) or more frequently, if
required (for example in case of outdoor pools)!
WARNING:
Please observe the detailed description for the setting of the
measuring cell. The initial pressure at the measuring cell should not exceed 1 bar.