EasyManua.ls Logo

ChemoformGroup CF Control 100 - Service; Basic Settings; Function: Delete Data Memory (Reset Factory Settings); Parameter Table for Service Staff

ChemoformGroup CF Control 100
60 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operating instructions
CF Control 100
Page
52
02/2017
15 Service / Basic settings
This chapter refers to settings which may only be carried out by qualified personnel
for initial commissioning and after fundamental changes to the system.
15.1 Function: Delete data memory (reset factory settings)
This function should only be used by trained service staff. This is why this function
cannot be easily selected and triggered. All settings are reset to the original factory
default parameters. The menu navigation and the following parameters are used for
a reset:
Parameters
Function
00
CODE
00 = End user code
NN = Code for pool dealers / service staff
NN: Code for dinotec factory service.
997: Load factory settings (excluding calibration values)
999: = Set calibration values (0mV, 58mV/pH)
15.2 Parameter table for service staff
Entry of service code C is required for access to the following parameters:
Parameters
Function
Facto
ry settings
8
Minimum flow in l/h (10
50)
15 l/h
19
Flow in pulse/l (0
-
1000)
0
51 Max. pulse frequency relay 1
in (pulse/h * 100)
72
(= 7200 pulse/h)
52 Period relay 1 in sec. 30 sec.
53
Minimum pulse duration relay 1 in 0.1
sec.
10 (=1 sec.)
56 Max. pulse frequency relay 2
in (pulse/h * 100)
57
Period relay 2 in sec.
30 sec.
58
Minimum pulse duration relay 2 in 0.1
sec.
10 (= 1 sec.)
Note: A re-calibration of the pH electrode is not required after having
entered the function value 997 (factory settings).

Table of Contents