Operating instructions
CF Control 100
15 Service / Basic settings
This chapter refers to settings which may only be carried out by qualified personnel
for initial commissioning and after fundamental changes to the system.
15.1 Function: Delete data memory (reset factory settings)
This function should only be used by trained service staff. This is why this function
cannot be easily selected and triggered. All settings are reset to the original factory
default parameters. The menu navigation and the following parameters are used for
a reset:
Parameters
00
CODE
00 = End user code
NN = Code for pool dealers / service staff
NN: Code for dinotec factory service.
997: Load factory settings (excluding calibration values)
999: = Set calibration values (0mV, 58mV/pH)
15.2 Parameter table for service staff
Entry of service code C is required for access to the following parameters:
51 Max. pulse frequency relay 1
72
52 Period relay 1 in sec. 30 sec.
53
Minimum pulse duration relay 1 in 0.1
10 (=1 sec.)
56 Max. pulse frequency relay 2
58
Minimum pulse duration relay 2 in 0.1
10 (= 1 sec.)
Note: A re-calibration of the pH electrode is not required after having
entered the function value 997 (factory settings).