EasyManua.ls Logo

ChemoformGroup CF Control 100 - Control of the Filter Pump; Setting the Time [Parameter 40]; Setting the Filter Times Switching Marks [Parameters 20 - 37]

ChemoformGroup CF Control 100
60 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operating instructions
CF Control 100
Page
38
02/2017
9 Control of the filter pump
9.1 Setting the time [parameter 40]
When you connect a filter pump for time-controlled operation, a time must be entered
first. In this example, the clock is to be set to 20:16.
Press the Eingabe / ENTER button and choose parameter P 40 to set the hour
value (confirm with ).
Set the value (right side) using or to the desired
hour, e.g 20:00 h.
Eingabe / ENTER
Select parameter 41 for the minute value and set the desired time on the right side.
In this example, the time is 20:16:
Eingabe / ENTER.
Press to return to the display level.
9.2 Setting the filter times switching marks [parameters 20 - 37]
You can switch a connected filter pump on and off punctual to the minute up to four
times a day. This is implemented by setting switching marks and assigning times to
these marks. The switching marks can be set to the conditions "On" , "Off"
 or "Inactive" . These switching marks are evaluated at the corresponding
times, and the pump is connected, if required.
The following is an example for setting and activating switching mark 1 to turn-on
time 20:15.
Press the Eingabe / ENTER button and choose parameter P 20 to set the
switching mark. (confirm with ).
Set the value (right side) using or to the desired
hour, e.g 20:00 h.
Note: If the CF Control 100 is disconnected (switched off) from the mains, the
clock is operated via battery.

Table of Contents