ECO 2 Ultrasound Diagnostic System
Page 27 / 189
Near heat generators
High humidity
NOTE:
This equipment generates, uses and can radiate radio frequency energy. The equipment may
cause radio frequency interference to other medical and non-medical devices and radio
communications. To provide reasonable protection against such interference, this product
complies with emissions limits for a Group 1, Class A Medical Devices Directive as stated in EN
60601-1-2. However, there is no guarantee that interference will not occur in a particular
installation.
If this equipment is found to cause interference (which may be determined by turning the
equipment on and off), the user (or qualified service personnel) should attempt to correct the
problem by one or more of the following measure (s):
Reorient or relocate the affected device(s).
Increase the separation between the equipment and the affected device.
Power the equipment from a source different from that of the affected device.
Consult the point of purchase or service representative for further suggestions.
3.5.1.2 Transport and Storage Environmental Requirement
The following environmental transport and storage conditions are within system tolerances:
Temperature: -5º C ~ 40º C
Relative Humidity: ≤ 80% non-condensing
Atmosphere Pressure: 700hPa ~ 1060hPa
3.5.1.3 Electrical Requirements
Power Consumption: less than 60 VA
Voltage Fluctuation
WARNING:
Maintain a fluctuation range of less than ±10% of voltage labeling on rear panel of the system,
otherwise the system may be damaged.
Grounding