ECO 2 Ultrasound Diagnostic System
Page 28 / 189
Before connecting the power cable, connect the attached ground protection cable from
Equipotentiality terminal on system rear panel to a specialized grounding device.
NOTE:
Please follow the outlined power requirements. Only use power cables that meet the system
guidelines—failure to follow these procedures may produce system damage.
Line power may vary in different geographic locations. Refer to the detailed ratings on the rear
panel of the system for detailed information.
Battery
To avoid the battery bursting, igniting, or fumes from the battery; causing equipment damage,
observe the following precautions: Do not immerse the battery in water or allow it to get wet. Do
not put the battery into a microwave oven or pressurized container. If the battery leaks or emits an
odor, remove it from all possible flammable sources. If the battery emits an odor or heat, is
deformed or discolored, or in a way appears abnormal during use, recharging or storage,
immediately remove it and stop using it. If you have any questions about the battery, short term
(less than one month) storage of battery pack: Store the battery in a temperature range between 0
degrees C (32 degrees F) and 50 degrees C (122 degrees F).
Long term (3 months or more) storage of battery pack: Store the battery in a temperature range
between-20 degrees C (-4 degrees F) and 45 degrees C (113 degrees F); Upon receipt of the
ECO and before first time usage, it is highly recommended that the customer performs one full
discharge/charge cycle. If the battery has not been used for >2 months, the customer is
recommended to perform one full discharge/charge cycle. It is also recommended to store the
battery in a shady and cool area with FCC (full current capacity).• One Full Discharge/Charge
Cycle Process:1. Full discharge of battery to let the ECO automatically shut down.2. Charge the
ECO to 100% FCC (full current capacity).3. Discharge of Venue 40 for complete shutdown(takes
one hour for discharge).• When storing packs for more than 6 months, charge the pack at least
once during the 6 month timeframe to prevent leakage and deterioration in performance.
3.5.1.4 Operation Space
Please leave enough free space from the back of the system to ensure well ventilation.
CAUTION: Leave enough free space from the back of the system; otherwise, with the
increasing of the temperature inside the unit, malfunction may occur.