ECO 2 Ultrasound Diagnostic System
Page 30 / 189
during transport. Any movement, coupled with the weight of the system, could cause it to break
loose.
Drive carefully to prevent damage from vibration. Avoid unpaved roads, excessive speeds, and
erratic stops or starts.
3.5.2 Powering the System
3.5.2.1 Acclimation Time
After being transported, the unit requires one hour for each 2.5 º increment if its temperature is
below 10 ºC or above 40 ºC.
NOTE:
Please keep at least 20 to 30 cm spare space away from the back of the system to ensure well
ventilation. Otherwise, with the increasing of the temperature inside the unit, malfunction may
occur.
3.5.2.2 Connecting the electric power
After making sure that the AC power supply in hospital is in normal status, and this AC voltage type
matches to the power requirements indicated on the label of system, then please connect the plug
of power cord to the DC IN socket at the rear panel of the system, and connect the other end of
power cord to the AC power supply socket in hospital.
Please use the power cable provided by the manufacturer, other type of power cable is not
allowed.
Press 1 s, system boot, Press and pop up dialog for shutting down. Click the enter
key then power off. Or press 3 s, system will be forced to shut down.
CAUTION:
Connecting the system to the wrong AC power supply may cause damage to the system and
danger to the operators and animals.
3.5.3 Probe Installation
CAUTION: Please only use the probes provided by manufacturer for this model, other
types of probes are not allowed to use with this system! Otherwise it may cause the damage to the