Remote Operation
5-21
*
CLS
*
ESE
*
ESE
?
*
ESR
?
*
IDN?
*
OPC
*
OPC
?
*RCL
*
RDT
?
*
RST
*SAV
*SRE
*
SRE
?
*STB
?
;
SP
<
NR
1
>
<
NR
1
>
<
NR1
>
<
NR
1
>
SP
SP
SP
*CLS Clear Status Command
Type: Device Status
Description: The *CLS command executes the following actions:
1. Clear these registers
<1> Channel Status Event registers for all channels
<2> Channel Summary Event register
<3> Questionable Status Event register
<4> Standard Event Status Event register
<5> Operation Status Event register
2. Clear the Error Queue
3. If “Clear Status Command” immediately follows a program
message terminator (<nl>), the “Output Queue” and the MAV bit
are also cleared.
Setting Syntax: *CLS
Setting Parameters: nil
*ESE Standard Event Status Enable Command/Query
Type: Device Status
Description: This command sets the condition of the Standard Event Status
Enable register to determine which event (see *ESR?) is allowed to
set the
ESB (Event Summary Bit) for the Status Byte register. A "1"
in the bit position enables the corresponding event. All of the events

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Chroma 63600 Series and is the answer not in the manual?

Chroma 63600 Series Specifications

General IconGeneral
Series63600
Operating Temperature0°C to 40°C
ModesCC, CV
InterfaceGPIB, USB
ProtectionOCP, OTP, OVP
Voltage Resolution1mV
CoolingForced Air