Query Example: CONF:DIO:OUT1?
CONFigure:DIO:OUT2
Type: Frame-Specific
Description: Set the DO2 type the pin No.43 in System I/O Port.
Setting Syntax: CONFigure:DIO:OUT2<space><NR1>
Setting Parameters: <NR1>, 0 ~ 7
0:NONE
1:OCP TEST PASS-H
2:OCP TEST FAIL-L
3:GONG TOTAL PASS-H
4:GONG TOTAL FAIL-L
5:OTP OVP OCP OPP REV-H
6:BUS CTRL. ACT-H
7:BUS CTRL. ACT-L
Setting Example: CONF:DIO:OUT2 1 Set DO2 to OCP TEST PASS-H.
CONF:DIO:OUT2 0 Set DO2 to NONE.
Query Syntax: CONFigure:DIO:OUT2?
Return Parameters: <NR1>, 0 ~ 7 [Unit = None]
Query Example: CONF:DIO:OUT2?
CONFigure:DIO:VOLTage
Type: Frame-Specific
Description: Set the voltage level of digital output pins (DO1&DO2).
Setting Syntax: CONFigure:DIO:VOLTage<space><NR1>
Setting Parameters: <NR1>, 0 ~ 3
0:NONE
1:1.8V
2:3.3V
3:5.0V
Setting Example: CONF:DIO:VOLT 2 Set output voltage is 3.3V.
CONF:DIO:VOLT 3 Set output voltage is 5.0V.
Query Syntax: CONFigure:DIO:VOLTage?
Return Parameters: <NR1>, 0 ~ 3 [Unit = None]
Query Example: CONF:DIO:VOLT?
DIO:OUT1
Type: Frame-Specific
Description: It sets t the system I/O port pin 9 DO1 status when the BUS CTRL.
mode is selected for DO1.
Setting Syntax: DIO:OUT1<space><NR1>
Setting Parameters: <CRD | NR1>, OFF(0), ON(1)
Setting Example: DIO:OUT1 ON Set DO1 to act.
DIO:OUT1 0 Set DO1 not to act.
Query Syntax: DIO:OUT1?
Return Parameters: <CRD>, OFF, ON [Unit = None]
Query Example: DIO:OUT1?
DIO:OUT2
Type: Frame-Specific
Description: It sets t the system I/O port pin 43 DO2 status when the BUS CTRL.
mode is selected for DO2.