Remote Operation
5-61
wave dynamic mode.
Query Syntax: FETCh:FREQuency?
Return Parameters:<NR2>, [Unit = Hertz]
Query Example: FETC:FREQ?
Return Example: 100.0
FETCh:POWer?
Type: Channel-Specific
Description: Returns the power measured at electronic load input.
Query Syntax: FETCh:POWer?
Return Parameters:<NR2>, [Unit = Watt]
Query Example: FETC:POW?
Return Example: 3.15
FETCh:STATus?
Type: Channel-Independent
Description: Returns real time status of the load module.
Bit Position
15~9
8
7
6
5
4
3
2
1
0
Condition
REMOTE
INHIBIT
FAN MAX LIM SYNC
REV
OPP
OC
P
OVP
OTP
Bit Weight
256
128
64
32
16
8
4
2
1
Query Syntax: FETCh:STATus?
Return Parameters:<NR1>, [Unit = None]
Query Example: FETC:STAT?
Return Example: 4
FETCh:TIME?
Type: Channel-Specific
Description: Returns the time measured in timing mode.
Query Syntax: FETCh:TIME?
Return Parameters:<NR2>, [Unit = Second]
Query Example: FETC:TIME?
Return Example: 0.045
FETCh:WH?
Type: Channel-Specific
Description: Returns the watt-hour measured in timing mode.
Query Syntax: FETCh:WH?
Return Parameters:<NR2>, [Unit = Watt-hour]
Query Example: FETC:WH?
Return Example: 20.045
FETCh:VOLTage?
Type: Channel-Specific
Description: Returns the voltage measured at electronic load input.
Query Syntax: FETCh:VOLTage?
Return Parameters:<NR2>, [Unit = Voltage]
Query Example: FETC:VOLT?
Return Example: 8.12

Table of Contents