Remote Operation
5-73
Arg5: <NR1>, 0 ~ 100, Resolution = 1, Unit = None.
Setting Example: PROG:DATA 1,STEP,2,0,5
Query Syntax: PROGram:DATA?<space><NR1>[<space><MAX | MIN>]
Return Parameters: <aard>
Query Example: PROG:DATA? 1
PROG:DATA? 1 MAX
PROG:DATA? 1 MIN
Return Example: 1,LIST,3,1,5
PROGram:DATA:LIST
Type: Channel-Specific
Description: Set the list parameters in program. (Note: All setting parameters in
this command can’t use suffix.)
Setting Syntax: PROGram:DATA:LIST<space><Arg1>,<Arg2>,<Arg3>,<Arg4>,
<Arg5>,<Arg6>,<Arg7>,<Arg8>,<Arg9>,<Arg10>,<Arg11>,
<Arg12>,<Arg13>,<Arg14>,<Arg15>,<Arg16>
Setting Parameters:
Selects a program to be set:
Arg1: <NR1>, 1 ~ 10, Resolution = 1, Unit = None.
Selects a sequence to be set:
Arg2: <NR1>, 1 ~ N, Resolution = 1, Unit = None.
Set the trigger mode of sequence:
Arg3: <NRf>, SKIP(0), AUTO(1), MANUAL(2), EXTERNAL(3),
Unit = None.
Set the run mode of sequence:
Arg4: <NRf>, CC(0), CR(1), CV(2), CP(3), Unit = None.
Set the mode’s range of sequence:
Arg5: <NRf>, LOW(0), MIDDLE(1), HIGH(2), Unit = None.
Set the load value according to run mode in sequence:
Arg6: <NRf>, Refer to respective specification for valid value
range.
Set the falling of slew rate in sequence:
Arg7: <NRf>, Refer to respective specification for valid value
range.
Set the rising of slew rate in sequence:
Arg8: <NRf>, Refer to respective specification for valid value
range.
Set the dwell time of sequence:
Arg9: <NRf>, 0.1ms ~ 30s, Resolution = 0.0001s, Unit =
Second.
Set the high-level of voltage specific in sequence:
Arg10: <NRf>, Refer to respective specification for valid value
range.
Set the low-level of voltage specific in sequence:
Arg11: <NRf>, Refer to respective specification for valid value
range.
Set the high-level of current specific in sequence:
Arg12: <NRf>, Refer to respective specification for valid value
range.
Set the low-level of current specific in sequence:
Arg13: <NRf>, Refer to respective specification for valid value
range.
Set the high-level of power specific in sequence:
Arg14: <NRf>, Refer to respective specification for valid value