Programmable DC Electronic Load 63600 Series Operation & Programming Manual
1-12
Model
63640-150-60
Configuration
400Wx1
Voltage *
1
*
8
0~150V
Current
0~1A
0~6A
0~60A
Power
90W
400W
400W
Static Mode
Typical min. operating voltage (DC)
0.3V@1A
0.3V@6A
1.8V@60A*15
Constant Current Mode
Range
0~1A
0~6A
0~60A
Resolution
0.02mA
0.1mA
1mA
Accuracy
0.04%+0.04%F.S.*17
Constant Resistance Mode
Range
CRL : 0.03~60Ω (400W/16V)
CRM: 0.64~800Ω (400W/80V)
CRH: 6.25~1.5kΩ (400W/150V)
Resolution *
9
1mS
Accuracy *
3
0.1%+0.067S (16V)
0.1%+0.00625S (80V)
0.1%+0.002S (150V)
Constant Voltage Mode
Range
16V
80V
150V
Resolution
1mV
1mV
10mV
Accuracy
0.025%+0.025%F.S.
Constant Power Mode
Range
0~8W
0~40W
0~400W
Resolution *
9
4mW
40mW
400mW
Accuracy *
4
0.3%+0.3%F.S.
Von/Voff Control*
13
Von&Voff Mode
CC / CR / CP
Accuracy
0.2%FS
Dynamic Mode - CC
Min. Operating Voltage *
11
*
16
1.8V
Frequency
100Hz~50kHz/0.01Hz~1kHz
Duty
1~99% (Min. Rise Time Dominated)
Accuracy
1µs/1ms+100ppm
Slew rate
0.2A/ms~
1.2A/ms~
12A/ms~
0.1A/µs
0.6A/µs
6A/µs
Resolution
0.02mA/µs
0.1mA/µs
1mA/µs
Accuracy
10% ±20µs
Min. Rise Time
10µs
Current
Range
0~1A
0~6A
0~60A
Resolution
0.02mA
0.1mA
1mA
Ext Wave
Mode
CC
Bandwidth
20kHz
Range
0~1A
0~6A
0~60A
Level
0~10V
Accuracy
0.5%F.S.
Program mode
Sequence No.
100/Program
Dwell / SEQ
0.1ms ~ 30s (Resolution : 0.1ms)
Load Setting
Refer to Static mode specifications
Spec Check
Voltage/Current/Power
Measurement
Voltage Read Back
Range
0~16V
0~80V
0~150V
Resolution
0.27mV
1.3mV
2.5mV
Accuracy *
5
0.025%+0.01%F.S.
Current read back
Range
0~1A
0~6A
0~60A
Resolution
0.02mA
0.1mA
1mA
Accuracy *
5
0.04%+0.04%F.S.
Power read back
Range
0~90W
0~400W
0~400W
Accuracy *
4
*
5
0.1%+0.1%F.S.
Voltage Monitor
Bandwidth
20 kHz

Table of Contents

Related product manuals