46
Cirris 4200 Series User Manual
The default setting of 0.010 seconds (10 milliseconds) should be adequate to test most cables. A longer duration will
increase the high voltage test time.
Since absorbed moisture may not have time to be dried out before the Insulation Resistance Test, consider using
the advaned Soak setting to dry out your insulation. A shorter soak time or not having soak time is more likely to
fail due to humidity related problems
Dielectric Withstand Duration
A longer duration provides a more stringent test for predicting dielectric failures (exhibited as sudden arcs); howev-
er, equivalent dielectric errors can be found by compensating a shorter duration with a higher high voltage setting.
A longer duration may be better for predicting some types of insulation failures and may increase the chance of
detecting a breakdown condition.