Calibration
49
25. Calibration
Every Cirris tester leaves the factory callibrated according to industry standards. Every two years, the tester will need to be
checked to ensure it still meets calibration requirements.
Cirris sells Performance Check Kits that run a series of tests to determine if your tester meets calibration standards. If the
tester passes the performance check, your tester is certied to run for another two years.
If your tester fails the performance check, speak to your Cirris sales rep. Cirris or a third party lab can calibrate your tester. No
external adjustments can be made to x the tester’s calibration.
Do not open the tester chassis or make any attempt to calibrate the tester yourself. This action will void your tester’s
warranty.