- 117 -
-
-
-
16.3
16.3
16.3
16.3 CO
CO
CO
CO
2
2
2
2
Menu
Menu
Menu
Menu
16.3.1
16.3.1
16.3.1
16.3.1 Parameter
Parameter
Parameter
Parameter Setup
Setup
Setup
Setup and
and
and
and Adjustment
Adjustment
Adjustment
Adjustment
Turn the knob to select and press CO
2
hot key on the screen to activate “ CO
2
Setup ” menu as shown
below:
Figure
Figure
Figure
Figure 1
1
1
1 6
6
6
6 -4
-4
-4
-4 CO
CO
CO
CO
2
2
2
2
Setup
Setup
Setup
Setup Menu
Menu
Menu
Menu
Following functions can be realized via CO
2
SETUP menu.
■ ALM: select “ ON ” to enable and store alarm prompt when CO
2
parameters have alarms. Select
“ OFF ” to disable alarm and display beside CO
2
. The default is “ ON ” .
■ ALM LEV: select from HIGH, MED and LOW. Level HIGH represents the most serious alarm,
followed by Level MED and Level LOW with a decrease of seriousness. Change in “ ALM LEV ” can
only affect the physiological alarm levels of CO
2
parameters including EtCO
2
upper limit, EtCO
2
lower limit, InsCO
2
upper limit, AwRR upper limit and AwRR lower limit. The default alarm level is
“ MED ” .
■
ALM REC: select “ ON ” to generate output from the recorder ever since CO
2
parameter alarm occurs.
The default is “ OFF ” .
■ CO
2
ALM HI: to adjust the upper alarm limit of EtCO
2
. If the measuring value is larger than CO
2
upper alarm limit, “ CO
2
TOO HIGH ” appears on the screen. After the measuring value returns to the
normal one, the information disappears.
■ CO
2
ALM LO: to adjust the lower alarm limit of EtCO
2
. If the measuring value is smaller than CO
2
lower alarm limit, “ CO
2
TOO LOW ” appears on the screen. After the measuring value returns to the
normal one, the information disappears.
■ INS ALM HI: to adjust the upper alarm limit of InsCO
2
. If the measuring value is larger than InsCO
2
upper alarm limit, “ INS TOO HIGH ” appears on the screen. After the measuring value returns to the
normal one, the information disappears.
■ AWRR ALM HI: to adjust the upper alarm limit of AwRR. If the measuring value is larger than the
upper alarm limit of AwRR, “ AWRR TOO HIGH ” appears on the screen. After the measuring value
returns to the normal one, the information disappears.
■ AWRR ALM LO: to adjust the lower alarm limit of AwRR. If the measuring value is smaller than the
lower alarm limit of AwRR, “ AWRR TOO LOW ” appears on the screen. After the measuring value
returns to the normal one, the information disappears.
■ UNIT: to change the display units of CO
2
and InsCO
2
parameters. “ mmHg ” and “ kPa ” are available
for selection.