60 Chapter 6 Installation Program
6.10 Troubleshooting for the Installation & Utilities Program
6.10.1 Serial Number Mismatch
Problem
If there is a mismatch between the serial number on the scanner label,
and the number in the scanner parameters (those saved in the scanner),
the program requests the correct serial number a second time.
In the following example, the second serial number digit, 3, is
incorrect, and should have been 9.
Problem Description and Solution
on page...
Serial Number Mismatch 60
Serial Number Mismatch (second time) 61
Wrong Media Inserted 63
SCSI Communication Failure 63
Diagnostics Check Failure 64
Calibration Slide Incorrectly Placed 65
Optics Check: Out of Tolerance 66