Do not use germicidal disposable wipes to clean the recessed lens.
Doing so may damage the lens and require handpiece replacement.
Debris from patients can accumulate on the lens. It may be
necessary
to clean the lens periodically during treatment.
39
EXCEL HR OPERATOR MANUAL
D1796, REV. D, 12/16
Clean the handpiece lens using a clean wood stick cotton swab and lens cleaning grade acetone, as
described below.
For optimal results, use acetone with 0.5% or lower water content,
which
is available at most camera shops. If unable to obtain acetone
with 0.5% or lower water content, contact Cutera Service (refer to
Appendix D).
Use only wood stick cotton swabs. Do not use products such as Q-
T
ips as they contain an adhesive that dissolves when it comes in
contact with acetone.
1. Ensure that the system is turned OFF or is in STANDBY mode.
2. Place a small amount of acetone on a wood stick cotton swab; if necessary
, shake the
excess acetone off of the tip prior to cleaning the handpiece lens.
3. Gently wipe the surface of the lens.
Recessed lens
CAUTION
NOTE
NOTE
NOTE