EasyManua.ls Logo

Dalsa Xtium-CLHS PX8 - Firmware Update: Manual Mode; Executing the Firmware Loader from the Start Menu; Figure 2: Manual Firmware Update

Default Icon
89 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Xtium-CLHS PX8 User's Manual Installing Xtium-CLHS PX8
13
Firmware Update: Manual Mode
Select Manual mode to load firmware other then the default version or when, in the case of
multiple Xtium-CLHS PX8 boards in the same system, if each requires different firmware.
The following figure shows the Device Manager manual firmware screen. Displayed is information
on all installed Xtium-CLHS PX8 boards, their serial numbers, and their firmware components.
Do a manual firmware update as follows:
Select the Xtium-CLHS PX8 to update via the board selection box (if there are multiple
boards in the system).
From the Configuration field drop menu select the firmware version required (typical
required or offered to support different CLHS cameras).
Click on the Start Update button.
Observe the firmware update progress in the message output window.
Close the Device manager program when the device reset complete message is shown.
Figure 2: Manual Firmware Update
Executing the Firmware Loader from the Start Menu
If required, the Xtium-CLHS PX8 Firmware Loader program is executed via the Windows Start Menu
shortcut Start • Programs • Teledyne DALSA • Xtium-CLHS PX8 Driver • Firmware
Update. A firmware change after installation would be required to select a different configuration
mode. See User Programmable Configurations.

Table of Contents

Related product manuals