TC1200 REFERENCE MANUAL
118
5
Keyboard Intercode Delay
This parameter specifies the intercode delay between the end of one code and the beginning
of the next. The delay can be set within a range of zero (0) to 99 seconds in 1 second
increments. A setting of zero specifies no delay.
To set the delay:
1. Determine the desired setting in seconds. (0-99).
2. Use the Chart in Appendix C to find the hex equivalent for the desired numeric value.
3. Go to page 43 and scan the ENT
ER/EXIT PROGRAMMING MODE barcode to enter
Programming Mode.
4. Scan the barcode: SELECT INTERCODE DELAY.
5. Scan the two hex equivalent characters from the keypad in Appendix B that represent the
numeric value determined in step 1 above.
NOTE
If you make a mistake before the last character, scan the CANCEL barcode
to abort and not save the entry string. You can then start again at the
beginning.
6. Scan the ENTER/EXIT PROGRAMMING MODE barcode to exit.
This completes the procedure.
5.2 READING PARAMETERS
5.2.1 Illumination Mode
Illumination Mode
Selects the illumination mode for the reader. The illumination mode can be used to:
to increase the overall diode life
to reduce power consumption;
to turn off the illuminator for safety purposes.
Selections:
Disabled: the illuminator is always off.
Enabled: the illuminator is always on.
Triggered: the illuminator is turned on when the reading phase starts. The illuminator is
turned off as soon as the code(s) are read or when the reading phase ends.
NOTE
The Triggered selection is ignored for Automatic and Test Operating
Modes.