30
Profile Probes Moisture Meter User Manual 4.0
Data
Select Options, Data to see this menu.
Plot ID
You can label each reading with a Plot ID.
Options - A to Z. Default: A.
Select Options, Data, Plot ID.
Scroll through the options, and press Set to select one.
Press Esc to return to the main menu.
Sample
A sample number is automatically assigned to each reading.
It automatically increments by one each time you store a reading.
Options: 1 to 2000. Default: 1.
Select Options, Data, Sample.
Select a number and press Set.
Press Esc to return to the main menu.
Device ID
You may label readings with a Device ID.
Options: 0 to 255. Default: 0
Select Options, Data, Device ID.
Select a number and press Set.
Press Esc to return to the main menu.
Root Depth
This is needed if you want soil moisture deficit (mmDef under Options, Display).
Options: 0 .. 9950 mm. Default: O
Select Options, Data Root Depth.
Select a number and press Set.
Press Esc to return to the main menu.
WARNING: Do not use the default settings to take mm Deficit readings. The default
Root Depth, set to zero, forces the calculated value for mm Deficit to be zero!
Some irrigation consultants recommend 700 mm as a suitable standard value for
Root Depth.