Electromagnetic immunity and emissions declaration
and guidance
The transmitter and receiver are intended for use in the electromagnetic environment
specified in the next table. The customer or the user of the transmitter should ensure
that it is used in such an environment.
Immunity test
Transmitter
compliance level
Receiver compliance level
Electrostatic
Discharge (ESD)
IEC 61000-4-2
± 8 kV Contact
± 15 kV Air
Magnetic Field (50Hz)
IEC 61000-4-8
30 A/m
Electrical Fast
Transient/Burst
IEC 61000-4-4
N/A ± 2 kV for power supply lines
Surge
IEC 61000-4-5
N/A ± 0.5 kV, ± 1 kV line(s) to line
(s)
160