E Technical Information | Dexcom G7 User Guide 161
Immunity test
Transmitter
compliance level
Receiver compliance level
Voltage Dips and
Interruptions
IEC 61000-4-11
IEC 60601-1-11
N/A 0% 230V for 1 cycle
0% 230V for 0.5 cycle at 8
phase angles
70% 230V (30% dip in 230V)
for 25 cycles
0% 230V for 250 cycles
Conducted Fields
Disturbance
IEC 61000-4-6
N/A 6 Vrms
150 kHz to 80 MHz
Radiated Fields
Disturbance
IEC 61000-4-3
10 V/m
at 80 MHz to 2700 MHz (AM Modulation)
Radiated and
Conducted Fields
Aircraft use
Meets RTCA /DO-160 edition G Section 20, Category T
Electromagnetic interference can still occur in the home health care environment as
control over the EMC environment cannot be guaranteed. An interference event can be
recognized by gaps in sensor readings or gross inaccuracies. The user is encouraged
to try to mitigate these effects by one of the following measures:
l
If your symptoms do not match your sensor readings, use your BG meter when
making treatment decisions. If sensor readings do not consistently match your
symptoms or BG meter values, then talk to your healthcare professional about