SELECTABLE OPERATING MODES
36
- After the button ENTER/PRINT is pressed: the indicator transmits on the PC serial the gross weight,
the tare, an ID code that uniquely identifies the weight; in addition the display show the message "
" for about 2 seconds.
The storage when pressing ENTER/PRINT follows the same conditions for the implementation of weighing
and printing, refer to paragraph “CONDITIONS FOR PERFORMING THE WEIGH AND PRINTS”.
The storage of the weigh through the PID serial command is always possible for all weights from 0 to full
scale, both with an approved instrument, both with an unapproved intrument; follows, therefore, only the
reenabling condition, refer to section “REENABLING THE PRINTOUTS”.
In case that these conditions are not observed:
- In response to the PID serial command it has “NO” instead of the ID.
- It has not no transmission in case of ENTER/PRINT key being pressed.
8.5.2 READING OF THE WEIGHS CARRIED OUT
In order to read the information relative to the weighs carried out:
- Press the MODE key.
- The message "" appears; now one should enter the rewriting number (from 00000 to 00255) and
press ENTER/PRINT.
- The message "" appears; now one should enter the weigh number (from 000000 to 131072) and press
ENTER/PRINT.
- Now it is possible to view on the display the weigh information in sequence, and scroll through it with
the ZERO key (ahead) or the TARE key (backwards):
"", in which is the scale number (always 1).
" " in which is the unit of measure (kg, g, t o lb).
gross weight (for about a second the message "" appears and then the gross weight value).
Tare weight (for about a second the message "", appears or "" if it is a manual tare,
then the tare value appears).
- Press C to return to weighing.
If the alibi memory is empty, when the MODE key is pressed the message “”appears for about a
second, an error acoustic signal is enabled and one returns to weighing.
If the entered ID is not valid, in other words, if there is no stored weigh relative to the entered ID, the
message " " appears and an error acoustic signal is enabled and one returns to weighing.
8.5.3 CANCELLATION OF THE ALIBI MEMORY
It is possible to delete all the weighs made, initializing the alibi memory; this operation is only possible in
the SETUP environment or via serial command ( ), with non approved instrument / sealed.