SELECTABLE OPERATING MODES
40
greater will the sampling be accurate.
With the 5-key indicator, setting the sampling time is only possible in the setup environment ( ).
8.7.4 PRINTING
If the presence of a printer has been configured and a sampling was performed, the execution of the weight
using the key ENTER/PRINT, will run a print showing also the percentage quantity present on the scale, in
addition to the programmed data.
8.8 DISPLAY WITH SENSITIVITY X 10 ()(TO BE USED IN TESTING DURING THE
CALIBRATION)
By pressing the MODE key one switches from the weight display with normal sensitivity to a sensitivity ten
times greater; in fact, one will note that the last digit on the right of the display will have a sensitivity equal
to the scale’s division divided by 10.
The printout can only be done when the indicator has the standard sensitivity.
In case the instrument is APPROVED, when “MODE” is pressed, the sensitivity times 10 is displayed for five
seconds after which the instrument returns to standard weight displaying.
In case the instrument is APPROVED for direct sale to the public, that view is only possible with capacities
greater than or equal to 100 kg (220 lb).
8.9 HOLD: FREEZING THE WEIGHT ON THE DISPLAY ()
By pressing MODE, the value of the weight is held on the display, and the display shows “”, alternately
with the held weight value (every 5 sec). To release the weight value on the display, press MODE key again
(the “” message is displayed).
By pressing MODE at length it is possible to enter in a menu for the selection of the type of functioning:
“”(static functioning, previously described), “”…””(dynamic functioning, for
weighing of moving objects, for example animal weighing; the higher the numeric parameter, greater is the
intervention of the HOLD function).
Press ZERO or TARE key to change the parameter and ENTER/PRINT to confirm. Then press MODE to
enable the selected functioning; by pressing again the MODE key the function is disabled.