3 • Operation and Maintenance
Doc. 065207-02 9/08 51
2. Move the cursor to METHOD EDITOR and press ENTER. The METHOD EDITOR
screen appears (see Figure 3-18
). All parameters do not fit on one screen. To
view additional parameters, press the down arrow button.
The
METHOD EDITOR screen displays the parameters for the method last
selected on the
SETUP screen (see Section C.1.3).
3. To select a different method to edit:
a. Press
ENTER to move the cursor to the method editing field.
b. Press an arrow button to step through the method numbers (1 through 24).
c. When the desired method number is displayed, press
ENTER.
4. To edit the method:
a. Press an arrow button to move the cursor to the first parameter that you
want to change (for example,
TEMPERATURE).
b. Press
ENTER. The cursor moves to the editing field.
NOTE If you decide not to edit this field, press MENU to
return the cursor to the left margin of the screen.
c. Press an arrow button to step through the values allowed for the
highlighted value. (Pressing and holding the arrow button steps through
the values more rapidly.) When the required value is displayed, press
ENTER.
d. If other changes are required, repeat Step a
through Step c.
Figure 3-18. Method Editor Screen (All Parameters Shown)
Press the down
arrow button to
scroll through the
list of parameters.
TEMPERATURE
METHOD
STATIC TIME
60%
5MIN
100C
1
RINSE VOLUME
PURGE TIME
100SEC
1
SST
STATIC CYCLE
CELL TYPE
1
SAVE