Reprocessing
Procedure:
1 Remove soiling immediately. Use a cloth dampened with disinfectant to remove
soiling. Do not spray cleaning solution directly onto the surface of the
device.
2 Perform surface disinfection.Thoroughly wipe the device 3 times. Observe the
recommended contact time specified by the disinfectant manufacturer.
3 After the product has been exposed to the disinfectant for the specified contact
time, remove residual disinfectant.
4 Wipe with a cloth dampened with water (preferably drinking-water quality). Allow
the product to dry.
5 Check the product for visible soiling. Repeat steps 1 to 5 if necessary.
6 Check the product for visible damage and replace if necessary.
12.5.4 Machine cleaning with thermal disinfection
Use a washer-disinfector that meets the requirements of the standard ISO 15883.
1 Securely position the product in the basket horizontally, in an open basket, or
directly on the tray. Ensure the following:
– All surfaces and interior spaces can be flushed completely.
– The water can drain off freely.
2 Use a suitable cleaning agent (refer to Validated reprocessing procedures on
page 138).
3 Select a suitable cycle (refer to Validated reprocessing procedures on page
138).
4 Use demineralized/deionized/distilled water for the final rinsing.
5 After the cycle has ended, check the product for visible soiling. If necessary,
repeat the cycle.
6 Check the product for visible damage and replace if necessary.
7 Allow the product to dry completely.
12.5.5 Additional reprocessing measures
Do not spray cleaning solution directly onto the surface of the device.
12.5.5.1 Reprocessing the shell, trolley, cabinet stand, and condensation
management bracket (if installed):
1 Wipe any fluids that may have accumulated in the heater well. Take special care
to clean and disinfect the heater well, the drain plug opening, and the drain plug
(on devices without the humidification system option).
CAUTION
Risk of injury or damage to the device
On devices with the condensation management system option, the hose/plug
assembly is disposable.
► Do not clean and reuse the hose/plug assembly.