Instructions for use Isolette 8000 plus SW 5.n 141
Reprocessing
2 Check for fluids that may have dripped onto the bottom surface of the water
reservoir opening. If fluids are present, use a clean paper towel dampened with
a cleaner/disinfectant to wipe the surfaces dry.
3 Follow surface disinfection recommendations to wipe with a cloth dampened
with disinfectant. Do not spray cleaning solution directly onto the surface of
the device.
12.5.5.2 Reprocessing the controller and the sensor module
1 When reprocessing the underside of the controller near the on/off switch, follow
surface disinfection recommendations to wipe with a cloth dampened with
disinfectant. Do not spray cleaning solution directly onto the surface of the
device.
2 Use a low-level or intermediate-level disinfectant to clean all surfaces
thoroughly, including the controller and sensor module. Refer to Disinfectants on
page 138.
3 Be sure to clean all holes and indentations; then dry with a clean cloth or paper
towel.
12.5.5.3 Reprocessing the hood, and inner double walls
NOTICE
Risk of damage to the device
Alcohol and Ultraviolet radiation can cause cracking and/or crazing (small stress
cracks) of the clear acrylic.
► Do not use alcohol for cleaning.
► Do not expose the clear acrylic to direct radiation from germicidal lights.
1 To release the inner double wall, press the catches (A) at the top of the inner
double wall.
P
The inner double walls are hinged on the access panels of the incubator.
2 Use a low-level or intermediate-level disinfectant to clean all surfaces of the
hood thoroughly, including the inner double walls, hand ports, and access
panels. Do not spray cleaning solution directly onto the surface of the
device. Refer to Disinfectants on page 138.
3 Be sure to clean all holes and indentations; then dry with a clean cloth or paper
towel.
072