EasyManua.ls Logo

Dräger Isolette 8000 plus - Page 64

Dräger Isolette 8000 plus
192 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
64 Instructions for use Isolette 8000 plus SW 5.n
Getting started
1 Unplug the power cable from the controller on the front of the incubator.
The Power Failure alarm sounds.
The Power Failure indicator on the front panel of the controller
illuminates.
2 Reconnect the power cable to the controller.
The incubator self-test runs.
Display 1 appears on the screen.
3 Check the Low Air Temperature alarm.
1 Rotate the locking knobs and open the front or rear access panel of the
incubator.
Within approximately 5 minutes, the Low Air Temperature alarm
message is displayed in the Trend/Alarm window.
The acoustic alarm signal sounds.
The alarm does not occur until the temperature falls 2.5 °C (4.5 °F) below the
set point. At high ambient temperatures, fanning the air within the hood can be
performed to induce the alarm.
2 Close the front or rear access panel, and rotate the locking knobs until they
are fully engaged.
4 Check the Low Skin Temperature alarm.
1 Set the air temperature to 35 °C (95 °F), and allow the temperature to
stabilize for 1 hour.
2 Insert a skin probe into the central skin probe connector of the sensor
module.
3 Place the skin probe 10 cm (4 in) above the center of the bed.
4 Set the skin set temperature to 35 °C (95 °F).
5 Rotate the locking knobs, and open the front or rear access panel.
Within approximately 5 minutes, the Low Skin Temperature alarm
message is displayed in the Trend/Alarm window.
The acoustic alarm signal sounds.
The alarm does not occur until the temperature falls 0.5 °C (0.9 °F) or 1.0 °C
(1.8 °F) (depending on the skin temperature alarm limit setting) below the set
point. At high ambient temperatures, fanning the air within the hood can be
performed to induce the alarm.
6 Close the front or rear access panel, and rotate both latches until fully
engaged.
7 Press the Audio Paused/Reset key.
The alarm is silenced.
5 Check the Central Probe Missing alarm.

Table of Contents

Other manuals for Dräger Isolette 8000 plus

Related product manuals