Instructions for use Isolette 8000 plus SW 5.n 65
Getting started
1 In the skin temperature mode, disconnect the skin probe from the central skin
probe connector of the sensor module.
The skin display goes blank.
The Central Probe Missing alarm message is displayed in the
Trend/Alarm window.
The acoustic alarm signal sounds.
2 Press the Audio Paused/Reset key.
The acoustic alarm signal silences for 5 minutes.
3 Reconnect the skin probe in the central skin probe connector.
The incubator returns to normal operation.
6 Check the maximum air temperature.
1 Connect a central skin probe to the sensor module, and select skin
temperature mode.
2 In the temperature override mode, select a temperature >37 °C (see Skin
temperature mode on page 79).
3 Place the skin probe outside the incubator.
4 Allow the incubator to heat.
5 If the High Temp CutOut alarm sounds, press the Audio Paused/Reset
key.
6 Verify that the incubator does not heat above 39.9 °C (103.82 °F) as
indicated on the screen.
7 Check the Connect Central Skin Probe alarm.
1 While operating in the air temperature mode, disconnect the central skin
probe.
2 Select the skin temperature mode.
The Connect Central Skin Probe alarm sounds.
8 Check the screen intensity feature at the system displays (see Keys on page
41).
Use the Up Arrow key to increase the screen intensity 3 levels.
Use the Down Arrow key to decrease the screen intensity 3 levels.
7.2.2 Hood/shell functional check
1 Check the access panels:
1 Rotate the locking knobs, and open the front access panel.
At the controller screen, the access panel open symbol ( ) appears.