Getting started
2 Pivot the access panel to the full open position (hanging straight down) (A).
3 Close the access panel, and rotate both latches until fully engaged.
At the controller screen, the access panel open symbol turns off.
4 Repeat steps 1 through 5 for the rear access panel.
CAUTION
Risk of injury
Both locking knobs must be fully engaged to avoid accidental opening of the
panels.
► Ensure that locking knobs are secure.
2 Check the hood operation:
1 If equipped with a weighing scale, first disconnect the weighing scale cable
as follows:
– Rotate the locking knobs and open the front or rear access panel.
– Disconnect the weighing scale cable from the sensor module.
– Close the access panel, and rotate both latches until they are fully
engaged.
2 Slowly tilt the hood back until the hood locks in place (B).