Instructions for use Isolette 8000 plus SW 5.n 79
Operation
8.1.1.1 Activating temperature override mode when setting the temperature
1 Press the Up Arrow key (E) and raise the set temperature to 37.0 °C (98.6 °F).
2 To activate the temperature override mode, press the >37°C button (F).
The >37°C LED indicator (G) illuminates and the color of the >37°C button
label changes.
The >37°C button is inoperative until the air set temperature is set to 37.0 °C
(98.6 °F).
3 To raise the set temperature beyond 37.0 °C (98.6 °F) up to 39.0 °C (102.2 °F),
press the Up Arrow key (E).
Pressing this key raises the temperature in 0.1° increments.
8.1.2 Skin temperature mode
The incubator must be pre-warmed for at least 1 hour in air temperature mode
before switching to skin temperature mode.
To select the set point in skin temperature mode:
1 If necessary, unlock the keypad.
If no keys are pressed within 15 seconds of making any selection, the screen
automatically reverts to the previous display.
053