82 Instructions for use Isolette 8000 plus SW 5.n
Operation
Dräger recommends that the central skin temperature and peripheral skin
temperature are measured and monitored. If 2 skin temperature probes are
used, changes in skin temperature can be detected more quickly using the
thermomonitoring function. Dräger recommends the use of original Dräger skin
temperature probes.
The following table shows the differences between the skin temperature probes:
1 While operating in the skin temperature mode, insert a yellow skin probe into the
central skin probe connector (A).
2 If monitoring 2 temperatures, insert a white skin probe into the peripheral skin
probe connector (B).
Symbols on
the device
Probe color Temperature measure-
ment
Attachment position
Yellow Central skin temperature In the liver or kidney
region
White Peripheral skin tempera-
ture
Foot or hand
037