86 Instructions for use Isolette 8000 plus SW 5.n
Operation
When using high levels of humidity, Dräger recommends using the optional
condensation management system.
8.2.1 Filling the water reservoir
CAUTION
Risk of injury or damage to the device
Sterile water is not an acceptable substitute for distilled water.
Use distilled water only (<10 ppm total dissolved solids).
CAUTION
Risk of injury or damage to the device
Spilled liquids could cause a slip hazard.
If any liquids spill or leak around the device, dry the area.
Mineral deposits may form on the evaporator, depending on the quality of the
distilled water.
The water level is visible through the front of the water reservoir. In addition, the
minimum (A), half-full (B), and maximum (C) water marks on the reservoir latch can
be used to assess the water level.
089
A
B
C