Instructions for use Isolette 8000 plus SW 5.n 87
Operation
1 Pull the latch forward and slide the reservoir assembly (D) halfway out from the
front of the incubator shell.
2 Open the cover (E) on the filling inlet by rotating it.
3 Pour distilled water into the reservoir until the water level is just below the
bottom of the filling inlet. Do not overfill. The water level should not exceed the
maximum mark on the front latch of the reservoir.
4 Close the cover on the filling inlet.
5 Slide the reservoir assembly fully into the shell.
6 Lock it in place by pushing in the latch.
CAUTION
Risk of injury or damage to the device
Using an incomplete water reservoir can prevent the system from operating
correctly.
Ensure that the reservoir lid is installed.
Ensure that the cover on the filling inlet is closed.
8.2.2 Selecting the humidity mode
1 Pre-warm the incubator in air temperature mode to the temperature prescribed
by the attending physician or according to Nursery Standing Orders.
Incubator is warmed.