Operation
8.3.2 Setting the oxygen set point
To set the oxygen set point in accordance with the attending physician
recommendations:
1 Activate the oxygen mode (see Turning the oxygen mode on or off on page 90).
If no keys are pressed within 15 seconds of making any selection, the screen
automatically reverts to the previous display.
2 Adjust the oxygen set point:
To raise the set point within a range of 21% to 65%, press the Up Arrow key
(D).
To lower the set point within a range of 65% to 21%, press the Down Arrow
key (E).
3 To acknowledge the oxygen control setting, press the Home button (F).
Device returns to display 1.
4 Press the Keypad Lock key (G).
✓ Keypad is locked.