EasyManua.ls Logo

Dräger Isolette 8000 plus - Page 91

Dräger Isolette 8000 plus
192 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Instructions for use Isolette 8000 plus SW 5.n 91
Operation
8.3.2 Setting the oxygen set point
To set the oxygen set point in accordance with the attending physician
recommendations:
1 Activate the oxygen mode (see Turning the oxygen mode on or off on page 90).
If no keys are pressed within 15 seconds of making any selection, the screen
automatically reverts to the previous display.
2 Adjust the oxygen set point:
To raise the set point within a range of 21% to 65%, press the Up Arrow key
(D).
To lower the set point within a range of 65% to 21%, press the Down Arrow
key (E).
3 To acknowledge the oxygen control setting, press the Home button (F).
Device returns to display 1.
4 Press the Keypad Lock key (G).
Keypad is locked.
051
40
35
30
25
20
D
E
F
G

Table of Contents

Other manuals for Dräger Isolette 8000 plus

Related product manuals