EasyManua.ls Logo

Elcon HiD 2000 Series - Hid 2037;2038

Default Icon
129 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6-10
6.2.8 HiD 2037/2038
a) connect the power supply at the termination board supply terminals; feed the supply voltage, in series with the
first multimeter A1 (current mode) and with the resistor box Rc, at the input terminals; connect the second
multimeter A2 (current mode), in series with a 250 resistor, at the output terminals; set the supply at 24 V and
switch it on; the green power LED must go on;
b) set the resistor box to 1.1 K; the output current must be within 19.50 and 20.50 mA; the input current must be
the same, with an error lower than 20 µA; the red fault LED must be off.
c) set the resistor box to 5.6 K; the output current must be within 3 and 5 mA; the input current must be the
same, with an error lower than 20 µA; the red fault LED must be off.
d) set the resistor box to 1.8 K; the input current must be within 11 and 13 mA; the output current must be the
same, with an error lower than 20 µA; the red fault LED must be off.
e) connect the multimeter A2 at the output terminals without the external resistor (output short-circuit); the red fault
LED must be on;
f) disconnect the multimeter A2 at the output terminals (output open circuit); the input current must be lower than
1.2 mA; the red fault LED must be on;

Table of Contents