Risk of injury
Pressurised system
− The device is pressurised during operation. Removing reagent tubes during
operation may lead to an explosive escape of chemicals or sample material
and can cause injuries.
• Never remove any reagent tubes during operation.
• Use the software in order to put the device into the mode “Replace
reagents” before performing any work on the device.
Risk of burns
Hot catalyst furnace
• The furnace temperature of the catalyst furnace can be up to 800°C
during operation and cause severe burns.
• Please wear heat-resistant protective gloves when handling with the
catalyst furnace.
• Before any maintenance work can be performed, please wait until
the furnace temperature is < 40
0
C.
• Check the furnace temperatures in the ELEMENTS software.
• Replace the filling of the catalyst furnace only after a sufficiently
long cooling time.
Weight:165 kg – The centre of gravity of the device is located on the right side.
Dimensions (WxHxD):56 x 78 x 64 cm