Safety
17
CAUTION
C13.0000
Risk of injury
Pressurised system
The device is pressurised during operation. Removing reagent tubes during
operation may lead to an explosive escape of chemicals or sample material
and can cause injuries.
Never remove any reagent tubes during operation.
Use the software in order to put the device into the mode “Replace
reagents” before performing any work on the device.
CAUTION
C14.0094
Risk of burns
Hot catalyst furnace
The furnace temperature of the catalyst furnace can be up to 800°C
during operation and cause severe burns.
Please wear heat-resistant protective gloves when handling with the
catalyst furnace.
Before any maintenance work can be performed, please wait until
the furnace temperature is < 40
0
C.
Check the furnace temperatures in the ELEMENTS software.
Replace the filling of the catalyst furnace only after a sufficiently
long cooling time.
Caution
C15.0107
Weight:165 kg The centre of gravity of the device is located on the right side.
Dimensions (WxHxD):56 x 78 x 64 cm

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ELTRA ELEMENTRAC ONHp2 and is the answer not in the manual?

ELTRA ELEMENTRAC ONHp2 Specifications

General IconGeneral
BrandELTRA
ModelELEMENTRAC ONHp2
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals