Maintenance
67
10.5.6 Filling the reagent tube of the CO converter
In order to replace the filling of the catalyst (see Section
(6) in 10.5 Filling of reagent tubes) and proceed as
follows:
1. Turn the main switch of the analyser to switch
position 1.
2. Before removing the reagent tube, please see
Removing and installing of reagent tubes.
3. Remove the used filling from the reagent tube.
4. Clean the reagent tube.
5. Dispose of the material in accordance with local
rules and regulations.
6. Please make sure that the reagent tube is
undamaged.
7. In the event that it is damaged or too much
contaminated, use a new reagent tube (Item
number 11064-3002).
8. Fill the reagent tube with 1.5-2 cm glass wool
(Item number 90331).
9. Fill the reagent tube with 20 cm contactors
reagent (Item number 90270).
10. Fill the reagent tube with 1.5-2 cm glass wool
(Item number 90331).
11. Before inserting the filled reagent tube, please see
Removing and installing reagent tubes“.
The filling of the reagent tube has now been replaced.
10.6 Replacing the O-rings
The O-rings are only replaced during maintenance or due to obvious wear.
The O-rings, which need to be replaced, are listed in Section Spare parts.
1,5-2cm
90331
1,5-2cm
90331
90270
20cm

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ELTRA ELEMENTRAC ONHp2 and is the answer not in the manual?

ELTRA ELEMENTRAC ONHp2 Specifications

General IconGeneral
BrandELTRA
ModelELEMENTRAC ONHp2
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals