Maintenance
63
10.5.2 Fill the reagent tube of the catalyst
In order to replace the filling of the catalyst (see Section
(1) in „10.5 Filling of reagent tubes“) and proceed as
follows:
1. Turn the main switch of the analyser to switch
position 1.
2. Before removing the reagent tube, please see
Removing and installing of reagent tubes„“.
3. Remove the used filling from the reagent tube.
4. Clean the reagent tube.
5. Dispose of the material in accordance with local
rules and regulations.
6. Please make sure that the reagent tube is
undamaged.
7. In the event that it is damaged or too much
contaminated, use a new reagent tube (Item
number 11064-3002).
8. Fill the reagent tube with 1.5-2 cm quartz wool
(Item number 90330).
9. Fill the reagent tube with 20 cm copper oxide
(Item number 90289).
10. Fill the reagent tube with 1.5-2 cm quartz wool
(Item number 90330).
11. Before inserting the filled reagent tube, please see
Removing and installing reagent tubes“.
The filling of the reagent tube has now been replaced.
The Elementrac OH-p does not have a heated
catalyst.
1,5-2cm
90330
1,5-2cm
90330
90289
20cm