4.3 Filling the reagent tubes
Risk of injury
Hazardous chemicals
− When working on the device, hazardous chemicals can cause fatal
poisoning or severe skin burns .
• Please observe the safety data sheet in terms of the substances used.
• Please wear your personal protective equipment.
• Never eat or drink within the immediate vicinity of chemicals.
Cutting injuries
Broken glass
− Sample tubes and reagent tubes are made of glass and can break. Broken
glass can cause cutting injuries.
• Please check reagent tubes and sample tubes for any damage before
use.
• Please wear protective gloves and goggles when handling with
reagent tubes and sample tubes.
• Replace damaged reagent tubes and sample tubes.
• Do not touch broken glass with unprotected hands.
Risk for eye injuries
Chemicals
− When chemicals are replaced, small particles of the chemicals can float in
the air and burn the eyes.
• Please always wear protective goggles when working with chemicals.
• Please observe the safety data sheets in terms of the chemicals used.
In order to fill the reagent tubes, please proceed as follows:
1. Fill the reagent tube of the dust trap.
2. Fill the reagent tube of the water trap.
3. Fill the reagent tube of the CO converter.
4. Fill the reagent tube for the gas processing of the carrier gas.
5. Fill the reagent tube of the gas processing for the
thermal conductivity detector.
6. Fill the reagent tube of the catalyst furnace.