Risk of injury
Pressurised system
− The device is pressurised during operation. Removing reagent tubes during
operation may lead to an explosive escape of chemicals or sample material
and can cause injuries.
• Never remove any reagent tubes during operation.
• Use the software in order to put the device into the mode “Replace
reagents” before performing any work on the device.
When providing the gas supply for the device, please ensure to observe the relevant information
on the gas inlet of the device. The permissible gases and pressures are clearly marked there.
Providing the gas supply for the device as follows:
1. Please make sure that the gas supply of the carrier gas is switched off.
2. Connect the carrier gas inlet on the analyser (please see „Rear side of the analyser “) with
the operator’s carrier gas supply. Use the supplied hose for this purpose.
Please observe the necessary purity and the pressure of the carrier gas (please see
technical data).
3. Repeat this procedure for all of the carrier gases you are using.
4. Connect the exhaust air outlet on the device (please see „Rear side of the analyser “) to
an appropriate ventilation system.
5. Connect the compressed air connection on the analyser (see „Rear side of the analyser
“) to the operator's compressed air supply. Please observe the pressure of the
compressed air (see technical data).
Die Gas supply is provided.
Please make sure that the gas and compressed air supply is properly connected. Due to the fact
that the hot sample carriers pose a potential fire hazard, the hoses should not be laid directly on
the laboratory table.