Maintenance
76
Reinstall the furnace lock assembly. Please make sure that the locking element
engages into the guiding element of the pneumatic cylinder.
Insert the two nuts and tighten them.
Put the upper furnace cover on from above and fasten it with the screw.
Close the window in the Elements software.
Turn the main switch of the analyser to position 2.
Perform a leak test in the software.
The upper electrode has been replaced.
10.7.4 Replace lower electrode (graphite tip)
CAUTION
C45.0000
Risk of injury
Moving parts
The furnace closes after user input. Hands in the area of the opening can
be crushed when the furnace closes.
Never reach into the oven when it is closing.
Use crucible tongs in order to place crucibles.
Turn the main switch of the analyser to position 1 or 0.
Loosen the 4 screws on the lower electrode.
Remove fitting and electrode.
Clean the surface and the fitting.