• Reinstall the furnace lock assembly. Please make sure that the locking element
engages into the guiding element of the pneumatic cylinder.
• Insert the two nuts and tighten them.
• Put the upper furnace cover on from above and fasten it with the screw.
• Close the window in the Elements software.
• Turn the main switch of the analyser to position 2.
• Perform a leak test in the software.
The upper electrode has been replaced.
10.7.4 Replace lower electrode (graphite tip)
Risk of injury
Moving parts
− The furnace closes after user input. Hands in the area of the opening can
be crushed when the furnace closes.
• Never reach into the oven when it is closing.
• Use crucible tongs in order to place crucibles.
• Turn the main switch of the analyser to position 1 or 0.
• Loosen the 4 screws on the lower electrode.
• Remove fitting and electrode.
• Clean the surface and the fitting.