Before replacing the filling of the gas processing, please
see items (2 and 3) in Fehler! Verweisquelle konnte n
icht gefunden werden.) and proceed as follows:
1. Turn the main switch of the analyser to switch
position 1.
2. Before removing the reagent tube, please see
„Removing and installing of reagent tubes“.
3. Remove the used filling from the reagent tube.
4. Clean the reagent tube.
5. Dispose of the material in accordance with local
rules and regulations.
6. Please make sure that the reagent tube is
undamaged.
7. In the event that it is damaged or too much
contaminated, use a new reagent tube (Item
number 88400-0006).
8. Fill the reagent tube with 1.5 to 2 cm glass wool
(Item number 90331).
9. Pour approx. 10 cm of magnesium perchlorate
(Item number 90200) onto the glass wool.
10. Pour 0.5 to 1 cm of glass wool onto the magnesium
perchlorate (Item number 90331).
11. Pour approx. 10 cm of sodium hydroxide (Item
number 90210) onto the glass wool.
12. Pour 1.5 to 2 cm of glass wool onto the sodium
hydroxide (Item number 90331).
13. Before inserting the filled reagent tube, please see
„Removing and installing reagent tubes“.
The filling of the reagent tube has now been replaced.