• Clean the seat of the flat seal.
• Clean the inside of the oven with a dry cloth.
• Insert the new electrode into the furnace from the bottom up. The electrode has a
coding to enable a better orientation. This must point to the right.
• Fasten the electrode with three new sealing washers and screws.
• Insert the new flat seal free of grease.
• Connect the compressed air hoses from the two vertical pneumatic cylinders. Please
make sure that the connection is correct: blue at the top; black on the bottom.
• Put the upper part of the furnace back on.
• Fasten the upper part of the furnace with the 4 screws.