Clean the seat of the flat seal.
Clean the inside of the oven with a dry cloth.
Insert the new electrode into the furnace from the bottom up. The electrode has a
coding to enable a better orientation. This must point to the right.
Fasten the electrode with three new sealing washers and screws.
Insert the new flat seal free of grease.
Connect the compressed air hoses from the two vertical pneumatic cylinders. Please
make sure that the connection is correct: blue at the top; black on the bottom.
Put the upper part of the furnace back on.
Fasten the upper part of the furnace with the 4 screws.