Maintenance
74
Loosen the four fastening screws of the upper part of the furnace and carefully remove
the furnace part.
Remove the compressed air hoses from the two vertical pneumatic cylinders.
Loosen the three screws which hold the electrode in place. Use the supplied Torx®
screwdriver for this purpose. Hold the electrode while performing this task.
NOTICE
Do not use a Torx® screwdriver with a ball head! The screws may be very tight and the
ball head could thereby tear off and damage the screw.
Remove the electrode of the furnace from below.
Remove the three screws with the sealing washers.
Remove the flat seal.