Maintenance
52
After 50 analyses or at least 2 times a day
Clean furnace and electrodes. Further information can be found in Section „Furnace
cleaning“.
After 500 analyses
Replace all chemicals (except copper oxide). Further information can be found in Section
„Removing and installing of reagent tubes“ and „Filling of reagent tubes“.
After 1000 analyses or by request
Replacing the crucible base made of graphite after 1000 analyses, or if the contact
surface between the crucible and the crucible base has significantly reduced due to
wear.„Replacing of electrodes“.
After 2000 analyses
Replace the copper oxide in the catalyst furnace.„Replacing the copper oxide in the
catalyst furnace“.
Take the electrodes out of the furnace and clean them with a brush. Further information
can be found in Section „Replacing the electrodes“.
NOTICE
The aforementioned maintenance recommendations are based on steel analysis and 99.995%
pure helium.
Depending on the application, these can differ significantly from the above recommendations.
For example, cleaning the furnace when analysing refractory metals (titanium) is recommended
after a maximum of 10 analyses. The service life of the copper oxide is also reduced the higher
the certain oxygen concentrations are (e.g. oxides, slags).
NOTICE
Set up a “counter” in the software. Please read the Section “Analyser Status” in the software
manual.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ELTRA ELEMENTRAC ONHp2 and is the answer not in the manual?

ELTRA ELEMENTRAC ONHp2 Specifications

General IconGeneral
BrandELTRA
ModelELEMENTRAC ONHp2
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals