Maintenance
59
CAUTION
C37.0000
Cutting injuries
Broken glass
Sample tubes and reagent tubes are made of glass and can break. Broken
glass can cause cutting injuries.
Please check reagent tubes and sample tubes for any damage before
use.
Please wear protective gloves and goggles when handling with
reagent tubes and sample tubes.
Replace damaged reagent tubes and sample tubes.
Do not touch broken glass with unprotected hands.
Activate the application for cooling down the catalyst furnace (Setting: Catalyst furnace
temperature = 0°C)
Open the front door of the analyser
Please wait until the catalyst furnace has cooled down.
Remove the cover of the catalyst furnace. For this purpose, loosen the screw on the
lower edge of the cover and remove the cover.
Screw
Cover