Maintenance
69
Remove the old O-ring and clean the surfaces.
Insert the new O-ring free of grease.
Reinstall the furnace lock assembly. Please make sure that the locking element
engages into the guiding element of the pneumatic cylinder.
Insert the two nuts and tighten them.
Put the upper furnace cover on from above and fasten it with the screw.
Close the window in the Elements software.
Turn the main switch of the analyser to position 2.
Perform a leak test in the software.
The seal of the upper furnace lock has been replaced.