Maintenance
72
.
NOTICE
Do not use a Torx® screwdriver with a ball head. The screws may be very tight and the
ball head could thereby tear off and damage the screw.
Remove the electrode of the furnace from below.
Clean the inside of the furnace with a dry cloth.
Insert the new electrode into the furnace from the bottom up. The electrode has a
coding to enable a better orientation. This must point to the right.
Fasten the electrode with the three screws